QUANTUM MAVERICK BALLOON CATHETER
Report
- Report Number
- 2134265-2010-04490
- Event Type
- Malfunction
- Date Received
- October 7, 2010
- Date of Event
- September 15, 2010
- Report Date
- September 15, 2010
- Manufacturer
- BOSTON SCIENTIFIC - MAPLE GROVE
- Product Code
- LOX
- PMA / PMN Number
- P860019
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4): IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4)
IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 99% STENOSED LESION WAS LOCATED IN A MODERATELY TORTUOUS LEFT ANTERIOR DESCENDING (LAD) CORONARY ARTERY. DURING THE FIRST INFLATION AT 14 ATMS THE 3.0 X 8MM QUANTUM MAVERICK MONORAIL BALLOON CATHETER RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | QUANTUM MAVERICK BALLOON CATHETER | CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS | LOX | BOSTON SCIENTIFIC - MAPLE GROVE | H7493808008300 | 11475040 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |