FDA Adverse Event Malfunction Summary report: N

QUANTUM MAVERICK BALLOON CATHETER

MDR report key: 1860754 · Received October 7, 2010

Report

Report Number
2134265-2010-04490
Event Type
Malfunction
Date Received
October 7, 2010
Date of Event
September 15, 2010
Report Date
September 15, 2010
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4)

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 99% STENOSED LESION WAS LOCATED IN A MODERATELY TORTUOUS LEFT ANTERIOR DESCENDING (LAD) CORONARY ARTERY. DURING THE FIRST INFLATION AT 14 ATMS THE 3.0 X 8MM QUANTUM MAVERICK MONORAIL BALLOON CATHETER RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUANTUM MAVERICK BALLOON CATHETER CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493808008300 11475040

Patients

Seq Age Sex Outcome Treatment
1