PHYSICA TT TIBIAL PLATE #5
Report
- Report Number
- 3008021110-2024-00008
- Event Type
- Injury
- Date Received
- January 30, 2024
- Date of Event
- December 19, 2023
- Report Date
- May 27, 2024
- Manufacturer
- LIMACORPORATE S.P.A.
- Product Code
- MBH
- PMA / PMN Number
- K201084
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
NO PRE-EXISTING ANOMALY WAS DETECTED BY THE CHECK OF THE STERILIZATION CHARTS OF THE LOT NUMBERS INVOLVED, THEREFORE WE CAN STATE THE DEVICES HAVE BEEN REGULARLY STERILIZED BEFORE BEING PLACED ON THE MARKET. NO PREVIOUS COMPLAINTS HAVE BEEN RECEIVED FOR THE LOT NUMBERS INVOLVED. ACCORDING TO OUR RECORDS, AT LEAST 29 OUT OF 30 WITH LOT. 2301067 - STER. 2300048 HAVE BEEN IMPLANTED AND THIS IS THE ONLY COMPLAINT RECEIVED ON THIS LOT NR. ACCORDING TO OUR RECORDS, AT LEAST 8 OUT OF 8 WITH LOT. 2222424 - STER. 2300023 HAVE BEEN IMPLANTED AND THIS IS THE ONLY COMPLAINT RECEIVED ON THIS LOT NR. ACCORDING TO OUR RECORDS, AT LEAST 17 OUT OF 18 WITH LOT. 2117937 - STER. 2100381HAVE BEEN IMPLANTED AND THIS IS THE ONLY COMPLAINT RECEIVED ON THIS LOT NR. ACCORDING TO OUR RECORDS, AT LEAST 66 OUT OF 72 WITH LOT. 21AT1AV - STER. 2100379 HAVE BEEN IMPLANTED AND THIS IS THE ONLY COMPLAINT RECEIVED ON THIS LOT NR. THE COMPLAINT SOURCE PROVIDED TO LIMACORPORATE TWO POST-OP X-RAYS AND PHOTOS FROM THE REVISION SURGERY. THE RECEIVED X-RAYS AND PHOTOS WERE SENT TO A MEDICAL EXPERT FOR EVALUATION. UNFORTUNATELY, THESE DOCUMENTS WERE NOT ENOUGH FOR THE MEDICAL EXPERT TO GIVE AN EVALUATION. THEREFORE, FURTHER X-RAYS (ESPECIALLY PRE-OPERATIVE) WERE REQUESTED TO COMPLAINT SOURCE, BUT THESE WERE NOT PROVIDED TO LIMACORPORATE. IN CONCLUSION, CONSIDERING THAT: 1) THE STERILIZATION CHARTS OF THE LOT NUMBERS INVOLVED IN THE EVENT HAVE BEEN CHECKED WITHOUT FINDING ANY PRE-EXISTING ANOMALY. 2) X-RAYS AND ADDITIONAL INFORMATION PROVIDED WERE NOT SUFFICIENT FOR THE MEDICAL EXPERT TO BE ANALYZED WE ARE NOT ABLE TO FURTHER INVESTIGATE THE EVENT, HOWEVER BASED ON THE FEW AVAILABLE INFORMATION RECEIVED WE CAN CLASSIFY THE EVENT AS NOT PRODUCT RELATED. PMS DATA: ACCORDING TO LIMACORPORATE PMS DATA, WE CAN ESTIMATE A REVISION RATE DUE TO INFECTION OF PHYSICA TT FIX. TIBIAL (FAMILY CODE: 6521.14.XXX), OF ABOUT 0,08%. NO CORRECTIVE ACTION NEEDED FOLLOWING THIS COMPLAINT, LIMACORPORATE WILL CONTINUE MONITORING THE MARKET TO PROMPTLY DETECT ANY FURTHER SIMILAR EVENT. THIS IS AN MDR FINAL REPORT.
NO PRE-EXISTING ANOMALY WAS DETECTED BY THE CHECK OF THE STERILIZATION CHARTS OF THE LOT NUMBERS INVOLVED, THEREFORE WE CAN STATE THE DEVICES HAVE BEEN REGULARLY STERILIZED BEFORE BEING PLACED ON THE MARKET. WE WILL SUBMIT A FINAL REPORT AFTER THE FINAL INVESTIGATION.
DURING A KNEE REVISION SURGERY PERFORMED ON (B)(6), 2023, DUE TO LOOSENING OF PHYSICA TT TIBIAL PLATE CAUSED BY INFECTION, THE ENTIRE PHYSICA PROTHESIS WAS REMOVED. A KNEE SPACER WAS PLACED, CEMENT WITH GENTAMICIN AND VANCOMYCIN. PRODUCTS EXPLANTED DURING SURGERY: PHYSICA PS - FEMORAL PEG (PRODUCT CODE: 6515.09.900, LOT. 2301067 - STER. 2300048). PS FEMORAL COMPONENT #5 LEFT (PRODUCT CODE: 6515.10.550, LOT. 2222424 - STER. 2300023). PHYSICA TT FIX. TIBIAL TRAY #5 (PRODUCT CODE: 6521.14.050, LOT. 2117937 - STER. 2100381). PHYSICA PS TIBIAL LINER #5 (PRODUCT CODE: 6535.50.510 , LOT. 21AT1AV - STER. 2100379). PREVIOUS SURGERY PERFORMED ON (B)(6), 2023. PATIENT IS A FEMALE, 65 YEARS OLD, HEIGHT: 160 CM , WEIGHT: 65 KG, HOUSEWIFE, EVENT OCCURRED IN ITALY.
KNEE REVISION SURGERY PERFORMED ON (B)(6) 2023, DUE TO LOOSENING OF PHYSICA TT TIBIAL PLATE CAUSED BY INFECTION. THE ENTIRE PHYSICA PROTHESIS WAS REMOVED AND G21 KNEE SPACER WAS PLACED, CEMENT WITH GENTAMICIN AND VANCOMYCIN. PRODUCTS EXPLANTED DURING SURGERY: PHYSICA PS - FEMORAL PEG (PRODUCT CODE: 6515.09.900, LOT. 2301067 - STER. 2300048) PS FEMORAL COMPONENT #5 LEFT (PRODUCT CODE: 6515.10.550, LOT. 2222424 - STER. 2300023) PHYSICA TT FIX. TIBIAL TRAY #5 (PRODUCT CODE: 6521.14.050, LOT. 2117937 - STER. 2100381) PHYSICA PS TIBIAL LINER #5 (PRODUCT CODE: 6535.50.510 , LOT. 21AT1AV - STER. 2100379) PREVIOUS SURGERY PERFORMED ON (B)(6) 2023. EVENT OCCURRED IN ITALY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2537522 | PHYSICA TT TIBIAL PLATE #5 | PHYSICA TT FIX. TIBIAL TRAY #5 | MBH | LIMACORPORATE S.P.A. | 6521.14.050 | 2117937 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Required Intervention |