INTELLIVUE MX750 PATIENT MONITOR
Report
- Report Number
- 9610816-2024-00046
- Event Type
- Malfunction
- Date Received
- January 30, 2024
- Date of Event
- January 7, 2024
- Manufacturer
- PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
- Product Code
- MHX
- UDI-DI
- 00884838083332
- PMA / PMN Number
- K210906
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
LOGS PROVIDED WERE REVIEWED AND ANALYZED BY PHILIPS PRODUCT SUPPORT ENGINEERING (PSE) AND IT WAS DETERMINED THAT THE PHILIPS EQUIPMENT WAS FUNCTIONING AS EXPECTED. PSE STATED THAT THE AUDIBLE ALARMING BEHAVIOR OF THE SPO2 SEARCHING INOP DID NOT CHANGE BETWEEN SW REV. L AND SW REV. N. THERE WAS NO TONE WITH REV L ON MX800 AND THERE IS NO TONE WITH REV N ON MX750.WITH THE MX750 SW REV. N, THE CUSTOMER IS USING X3 WITH MASIMO RAINBOW SET MEASUREMENT AND MASIMO SENSORS. THE MASIMO RAINBOW SET ALGORITHM IS OWNED BY MASIMO CORPORATION, AND SO IS THE MASIMO SENSOR TECHNOLOGY. PHILIPS CAN ONLY ASSIST IN DETAIL ON PHILIPS OWNED TECHNOLOGY. BASED ON THE INFORMATION PROVIDED IN THE CASE AND THE RESULT OF THE LOG EVALUATION PROVIDED BY THE PSE, THE CUSTOMER'S ALLEGATION COULD NOT BE CONFIRMED. THE DEVICE WAS CONFIRMED TO BE OPERATING PER SPECIFICATIONS AND NO FAILURE WAS IDENTIFIED.
E1: REPORTER INSTITUTION PHONE NUMBER : (B)(6). E1: REPORTER PHONE #: (B)(6). A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.
IT WAS REPORTED THAT THERE WAS NO AUDIBLE ALARMING ON MX750. ONLY VISUAL INOP IS DISPLAYED FOR SPO2 LOOSES SIGNAL. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
IT WAS REPORTED THAT THERE WAS NO AUDIBLE ALARMING ON MX750. ONLY VISUAL INOP IS DISPLAYED FOR SPO2 LOOSES SIGNAL. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1252072 | INTELLIVUE MX750 PATIENT MONITOR | INTELLIVUE MX750 PATIENT MONITOR | MHX | PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH | 866471 | 00884838083332 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |