FDA Adverse Event Malfunction Summary report: N

INTELLIVUE MX750 PATIENT MONITOR

MDR report key: 18607243 · Received January 30, 2024

Report

Report Number
9610816-2024-00046
Event Type
Malfunction
Date Received
January 30, 2024
Date of Event
January 7, 2024
Manufacturer
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
Product Code
MHX
UDI-DI
00884838083332
PMA / PMN Number
K210906
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

LOGS PROVIDED WERE REVIEWED AND ANALYZED BY PHILIPS PRODUCT SUPPORT ENGINEERING (PSE) AND IT WAS DETERMINED THAT THE PHILIPS EQUIPMENT WAS FUNCTIONING AS EXPECTED. PSE STATED THAT THE AUDIBLE ALARMING BEHAVIOR OF THE SPO2 SEARCHING INOP DID NOT CHANGE BETWEEN SW REV. L AND SW REV. N. THERE WAS NO TONE WITH REV L ON MX800 AND THERE IS NO TONE WITH REV N ON MX750.WITH THE MX750 SW REV. N, THE CUSTOMER IS USING X3 WITH MASIMO RAINBOW SET MEASUREMENT AND MASIMO SENSORS. THE MASIMO RAINBOW SET ALGORITHM IS OWNED BY MASIMO CORPORATION, AND SO IS THE MASIMO SENSOR TECHNOLOGY. PHILIPS CAN ONLY ASSIST IN DETAIL ON PHILIPS OWNED TECHNOLOGY. BASED ON THE INFORMATION PROVIDED IN THE CASE AND THE RESULT OF THE LOG EVALUATION PROVIDED BY THE PSE, THE CUSTOMER'S ALLEGATION COULD NOT BE CONFIRMED. THE DEVICE WAS CONFIRMED TO BE OPERATING PER SPECIFICATIONS AND NO FAILURE WAS IDENTIFIED.

Additional Manufacturer Narrative · 0

E1: REPORTER INSTITUTION PHONE NUMBER : (B)(6). E1: REPORTER PHONE #: (B)(6). A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS NO AUDIBLE ALARMING ON MX750. ONLY VISUAL INOP IS DISPLAYED FOR SPO2 LOOSES SIGNAL. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS NO AUDIBLE ALARMING ON MX750. ONLY VISUAL INOP IS DISPLAYED FOR SPO2 LOOSES SIGNAL. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1252072 INTELLIVUE MX750 PATIENT MONITOR INTELLIVUE MX750 PATIENT MONITOR MHX PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH 866471 00884838083332

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown