FDA Adverse Event Malfunction Summary report: N

ECHELON*FLEX60 LONG

MDR report key: 1860690 · Received October 7, 2010

Report

Report Number
3005075853-2010-05769
Event Type
Malfunction
Date Received
October 7, 2010
Date of Event
June 22, 2010
Report Date
July 7, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K081146
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): ARTICULATION LINK. THE ANALYSIS FOUND THAT ONE LONG60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO CARTRIDGE RELOAD PRESENT. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. UPON FURTHER INSPECTION THE JOINT COVER WAS NOTED TO BE TORN; THERE IS NO EVIDENCE THAT THERE IS ANY PORTION OF THE JOINT COVER MISSING. IN ADDITION, THE ARTICULATION LINK WAS FOUND BROKEN. WHILE NO CONCLUSION COULD BE REACHED AS TO HOW THE ARTICULATION LINK BECAME DAMAGED, IT SHOULD BE NOTED THAT A 100% INSPECTIONS TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A GASTRIC BYPASS PROCEDURE, THE DEVICE MISFIRED STAPLES. NO OTHER DETAILS WERE PROVIDED. NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECHELON*FLEX60 LONG STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK G4RR5X

Patients

Seq Age Sex Outcome Treatment
1