OT ULTRALINK METER
Report
- Report Number
- 2939301-2010-08906
- Event Type
- Malfunction
- Date Received
- October 7, 2010
- Report Date
- October 5, 2010
- Manufacturer
- LIFESCAN, INC.
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW UP # 1/SUPPLEMENTAL REPORT TEXT 12/06/2010. THE LAY USER/PATIENT'S METER HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE METER INVOLVED IN THIS CASE HAS PASSED TESTING WITH NO FAULTS FOUND. IF ANY ADDITIONAL INFORMATION IS AVAILABLE, THE FDA WILL BE NOTIFIED IN A SECOND FOLLOW UP REPORT. AT THIS TIME, WE CONSIDER THIS MATTER CLOSED.
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT; 510(K)# IS K073231.
A PATIENT REPORTED BLOOD GLUCOSE RESULTS OF "30 MG/DL", "111 MG/DL", "36 MG/DL" AND "203 MG/DL" WITH A LIFESCAN METER, PERFORMED WITHIN 20 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS THE EXPECTED VALUE OF <= 20% AND/OR <= 20 MG/DL. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WERE NO ALLEGATIONS OF HARM OR INJURY. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE DEVICE LOCKED ON THE BILE DUCT THEN IT WAS REMOVED BY RIPPING IT OFF FROM TISSUE. THERE WAS TISSUE DAMAGE THAT WAS REPAIRED BY CLIPPING AROUND THE TEAR. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT ULTRALINK METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN, INC. | 3035389 |