ALTRUA
Report
- Report Number
- 2124215-2010-15020
- Event Type
- Injury
- Date Received
- October 7, 2010
- Date of Event
- July 22, 2010
- Report Date
- April 19, 2017
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- PMA / PMN Number
- D970003
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THIS DEVICE REMAINS IMPLANTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.
THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THE PATIENT WITH THIS DEVICE EXPERIENCED A SYNCOPAL EPISODE.
THIS DEVICE WAS EXPLANTED APPROXIMATELY SEVEN YEARS LATER AND WAS RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ALTRUA | IMPLANTABLE PULSE GENERATOR | LWP | GUIDANT CRM CLONMEL IRELAND | S203 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Life Threatening | 4054| 1276| S203| 4055 |