ENDOWRIST
Report
- Report Number
- 2955842-2024-10814
- Event Type
- Malfunction
- Date Received
- January 30, 2024
- Date of Event
- January 8, 2024
- Report Date
- January 8, 2024
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874119815
- PMA / PMN Number
- K214095
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- 003
Narratives
INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT TO PERFORM FAILURE ANALYSIS. THE MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT WAS ANALYZED AND FOUND TO HAVE A BROKEN GRIP CABLE AT THE DISTAL END. THE COMPLAINT WAS CONFIRMED BY FAILURE ANALYSIS.
INTUITIVE SURGICAL, INC. (ISI) HAS NOT RECEIVED THE MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT FOR EVALUATION. THEREFORE, THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE HAS NOT BEEN DETERMINED. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE PRODUCT IS RETURNED AND EVALUATED AND/ OR IF ADDITIONAL INFORMATION IS RECEIVED.
REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT HAD BROKEN AND EXPOSED WIRES. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE CUSTOMER NOTED THE WIRE WAS EXPOSED ON INSTRUMENT. THERE WERE PROTRUDING CABLE THAT CONTROLS OPENING/CLOSING OF THE JAWS AND UP/DOWN MOTION OF THE WRIST.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2538289 | ENDOWRIST | MEGA SUTURECUT ND | NAY | INTUITIVE SURGICAL, INC | 471309-16 | K11230302 0108 | 00886874119815 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES |