FDA Adverse Event
Malfunction
Summary report: N
WARM TOUCH II COMPACT WARMING UNIT
MDR report key: 1860365
·
Received September 24, 2010
Report
- Report Number
- 2936999-2010-01215
- Event Type
- Malfunction
- Date Received
- September 24, 2010
- Date of Event
- August 1, 2010
- Report Date
- August 27, 2010
- Manufacturer
- COVIDIEN, FORMERLY TYCO HEALTHCARE
- Product Code
- DWJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE COVIDIEN SERVICE TECH PERFORMED INSPECTION AND PERFORMANCE TESTING AND FOUND THE TURBINE (FAN) WAS MISSING FROM THE UNIT, HOWEVER, THERE WAS NO EVIDENCE OF SMOKE OR ANY BURNED HOUSING OR COMPONENTS. IT COULD NOT BE CONFIRMED WHEN OR BY WHOM THE TURBINE WAS REMOVED. PART # 501-5800 IS NOT DISTRIBUTED IN THE U.S; HOWEVER THERE IS A DEVICE OF ESSENTIALLY IDENTICAL DESIGN DISTRIBUTED IN THE UNITED STATES. APPLICABLE 510K# FOR US DISTRIBUTED PART IS K913016.
Description of Event or Problem · 1
THE CUSTOMER REPORTED "THE UNIT WAS VERY NOISY AND SMELLS LIKE BURNED". NO PT INVOLVEMENT WAS REPORTED. THE COVIDIEN SERVICE TECH CALLED AND SPOKE TO THE CUSTOMER'S SERVICE TECH WHO STATED THE BLOWER WAS DEFECTIVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | WARM TOUCH II COMPACT WARMING UNIT | CONVECTIVE AIR WARMER | DWJ | COVIDIEN, FORMERLY TYCO HEALTHCARE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |