FDA Adverse Event Malfunction Summary report: N

BD MULTITEST¿

MDR report key: 18602795 · Received January 29, 2024

Report

Report Number
2647876-2024-00016
Event Type
Malfunction
Date Received
January 29, 2024
Date of Event
January 11, 2024
Report Date
July 11, 2024
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
GKZ
UDI-DI
00382906629678
PMA / PMN Number
K090967
Removal / Correction Number
RES# 94830
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REMEDIAL ACTION FIELD HAS BEEN UPDATED WITH CORRECTED INFORMATION: H.9 - REMEDIAL ACTION # - RES# 94830.

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY: BASED ON THE INVESTIGATION RESULTS, THE REPORTED ISSUE (NON-SPECIFIC STAINING CREATING UNUSUAL STAINING PATTERNS FOR CD4+CD+8 POPULATIONS) WAS CONFIRMED. INVESTIGATION RESULTS THAT WERE PERFORMED ON THE INDICATED FAILURE MODE WERE THE FOLLOWING: DATA PROVIDED BY CUSTOMER WAS REVIEWED. RETAIN TESTING RESULTS SHOWED THE PRESENCE OF A NON-SPECIFIC STAINING. POTENTIAL CAUSE IS DETERMINED TO BE MANUFACTURING RELATED. ADDITIONAL CONTROLS HAVE BEEN PUT IN PLACE IN THE MANUFACTURING STAGE SPECIFIC TO THE SUPPLIED COMPONENT. CORRECTIVE ACTION WAS INITIATED TO ADDRESS THE ROOT CAUSE AND PREVENTIVE ACTIONS. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

INITIAL REPORTER FACILITY NAME: (B)(6) HOSPITAL. G5. PMA/510K: K170974. H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD MULTITEST CONTAMINATION OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ANTIBODIES CONTAMINATED WITH SUBSTANCE. THE CUSTOMER FOUND THAT AS LONG AS THIS BATCH OF LOT IS STAINED, IN THE CD45 VS SSC , CD4 VS CD8 PLOTS, THERE ARE HUGE NOISE, BUT OTHER LOTS ARE NORMAL WITHOUT ANY NOISE. VERY CLEAN CD45 STAIN UNLIKE LOT 49506.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD MULTITEST CONTAMINATION OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ANTIBODIES CONTAMINATED WITH SUBSTANCE. THE CUSTOMER FOUND THAT AS LONG AS THIS BATCH OF LOT IS STAINED, IN THE CD45 VS SSC , CD4 VS CD8 PLOTS, THERE ARE HUGE NOISE, BUT OTHER LOTS ARE NORMAL WITHOUT ANY NOISE. VERY CLEAN CD45 STAIN UNLIKE LOT 49506.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD MULTITEST CONTAMINATION OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ANTIBODIES CONTAMINATED WITH SUBSTANCE. THE CUSTOMER FOUND THAT AS LONG AS THIS BATCH OF LOT IS STAINED, IN THE CD45 VS SSC , CD4 VS CD8 PLOTS, THERE ARE HUGE NOISE, BUT OTHER LOTS ARE NORMAL WITHOUT ANY NOISE. VERY CLEAN CD45 STAIN UNLIKE LOT 49506.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2538211 BD MULTITEST¿ COUNTER, DIFFERENTIAL CELL GKZ BECTON DICKINSON CARIBE LTD. 49506 00382906629678

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown