ACCU-CHEK ® PERFORMA TEST STRIPS
Report
- Report Number
- 1823260-2010-05914
- Event Type
- Malfunction
- Date Received
- October 7, 2010
- Date of Event
- July 30, 2010
- Report Date
- October 27, 2010
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- LFR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE EVENT OCCURRED IN (B)(6). TWO METERS WERE REPORTED TO HAVE BEEN IN USE WHEN THE REPORTED COMPARISONS WERE OBTAINED. CALLER DID NOT SPECIFY WHICH METER PRODUCED WHICH RESULT.
PATIENT IS MALE WITH COMMUNICATION ARTERY ANEURYSM, (B)(6). WHEN THE PHYSICIAN INSERTED THE FIRST STENT (ENC452812) VIA Y VALVE, THE STENT SYSTEM WAS ADVANCED TO MICROCATHETER HUB POSITION, IT WAS DIFFICULT TO ADVANCE, SO THE PHYSICIAN WITHDREW THE STENT, DURING THIS PERIOD, THE STENT WAS RELEASED. THEREFORE, THE PHYSICIAN CHANGED THE SECOND ONE (ENC453712) TO IMPLANT VIA MICROCATHETER, BUT IT WAS BLOCKED IN THE PROXIMAL SECTION OF THE MICROCATHETER AND CANNOT BE WITHDRAWN. THEN THE PHYSICIAN WITHDREW THE MICROCATHETER WITH THE STENT AND CHANGED ANOTHER MICROCATHETER TO COMPLETE THE PROCEDURE. BOTH THE ENTERPRISE INTRODUCERS WERE COMPLETELY SEATED IN THE MICROCATHETER HUB, BUT THE PHYSICIAN SAID THE STENT WAS DIFFERENT FROM OTHERS, IT WAS DIFFICULT TO ADVANCE. DURING INSERTION, THE TUOHY OR Y CONNECTOR WAS CLOSED TO SECURE THE ENTERPRISE INTRODUCERS AND PREVENT MOVEMENT. NO DAMAGES WERE NOTICED ON THE MICROCATHETER THAT MAY HAVE CONTRIBUTED TO THE REPORTED EVENT. PRIOR TO THE EVENT, NO OTHER PRODUCTS WERE ABLE TO GO ALL THE WAY THROUGH THE MICROCATHETER, HOWEVER THE AFTER THE EVENT, THE SAME MICROCATHETER WAS UTILIZED TO COMPLETE THE PROCEDURE. NEITHER, THE MICROCATHETER NOR THE ENTERPRISE DELIVERY SYSTEM DISTAL TIPS WERE RE-SHAPED. A CONSTANT FLUSH WAS MAINTAINED AT ALL TIMES THROUGH ALL THE SYSTEMS. AFTER REMOVAL FROM THE PATIENT, THE DEVICES (ENTERPRISE DELIVERY SYSTEM (DISTAL TIP (UNRAVELED, STRETCHED, KINK, BEND, FRACTURE, ETC), STENT (STRUT UPLIFT OR DEFORMED, ETC), OR MICROCATHETER) WERE DAMAGED. THE ANEURYSM WAS 8MMX6MM.
CALLER STATES NEONATE PATIENT RECEIVED THE FOLLOWING PERFORMA SYSTEM/LAB RESULTS WITHIN 10 MINUTES: 1) 51 MG/DL/34 MG/DL 2) 58 MG/DL/44 MG/DL 3) 55 MG/DL/40 MG/DL 4) 49 MG/DL/35 MG/DL 5) 55 MG/DL/37 MG/DL 6) 64 MG/DL/44 MG/DL 7) 63 MG/DL/44 MG/DL 8) 88 MG/DL/63 MG/DL 9) 78 MG/DL/53 MG/DL NO ACTIONS TAKEN BASED ON DEVICE RESULTS. NO ADVERSE EVENT REPORTED. REQUESTED RETURN OF SUSPECT DEVICE AND REPLACEMENT WAS SENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCU-CHEK ® PERFORMA TEST STRIPS | BLOOD GLUCOSE MONITORING TEST STRIPS | LFR | ROCHE DIAGNOSTICS | 370196 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |