FDA Adverse Event Malfunction Summary report: N

REPAIR CORE IMPACTION DRILL

MDR report key: 1860228 · Received September 21, 2010

Report

Report Number
1811755-2010-01221
Event Type
Malfunction
Date Received
September 21, 2010
Date of Event
August 25, 2010
Report Date
August 25, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HANDPIECE HAS BEEN RECEIVED AT THE MFR FOR INVESTIGATION. AN EVAL WAS CONDUCTED AND THE COMPLAINT WAS CONFIRMED. ACCORDING TO THE INVESTIGATION DETAILS, THE DRIVESHAFT BEARINGS WERE OUT OF BALANCE. THE ROTOR, DRIVESHAFT BEARINGS, AND NOSE CONE WERE REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HANDPIECE BEGAN OVERHEATING DURING A SURGICAL PROCEDURE. THERE WAS NO REPORTED PT OR USER INJURY, AND THE CASE WAS COMPLETED SUCCESSFULLY WITH ANOTHER DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REPAIR CORE IMPACTION DRILL BONE CUTTING INSTRUMENT AND ACCESSORIES DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK