FDA Adverse Event
Malfunction
Summary report: N
REPAIR CORE IMPACTION DRILL
MDR report key: 1860228
·
Received September 21, 2010
Report
- Report Number
- 1811755-2010-01221
- Event Type
- Malfunction
- Date Received
- September 21, 2010
- Date of Event
- August 25, 2010
- Report Date
- August 25, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HANDPIECE HAS BEEN RECEIVED AT THE MFR FOR INVESTIGATION. AN EVAL WAS CONDUCTED AND THE COMPLAINT WAS CONFIRMED. ACCORDING TO THE INVESTIGATION DETAILS, THE DRIVESHAFT BEARINGS WERE OUT OF BALANCE. THE ROTOR, DRIVESHAFT BEARINGS, AND NOSE CONE WERE REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE HANDPIECE BEGAN OVERHEATING DURING A SURGICAL PROCEDURE. THERE WAS NO REPORTED PT OR USER INJURY, AND THE CASE WAS COMPLETED SUCCESSFULLY WITH ANOTHER DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | REPAIR CORE IMPACTION DRILL | BONE CUTTING INSTRUMENT AND ACCESSORIES | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |