FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 1860160 · Received September 21, 2010

Report

Report Number
1720753-2010-03154
Event Type
Malfunction
Date Received
September 21, 2010
Date of Event
September 15, 2010
Report Date
September 21, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE HAND SWITCH, FOOT SWITCH AND GENERATOR INTERFACE BOARD. PERFORMED TOUCH SCREEN CALIBRATION. TESTED, SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

CUSTOMER REPORTED THE SYSTEM INTERMITTENTLY DISPLAYS X-RAY SECURITY ERROR AND THE TOUCH SCREEN IS NOT ACCURATE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1