FDA Adverse Event Malfunction Summary report: N

COOL PATH CATHETER 7F, 1304-CP-7-25-L

MDR report key: 1860133 · Received September 21, 2010

Report

Report Number
2030404-2010-00144
Event Type
Malfunction
Date Received
September 21, 2010
Date of Event
August 26, 2010
Report Date
August 26, 2010
Manufacturer
ST JUDE MEDICAL, IRVINE
Product Code
OAD
PMA / PMN Number
P060019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATIONS HAVE CONFIRMED A LEAK IN THE HANDLE OF THE CATHETER IS CAUSED BY THE LUMEN BREAKING AT THE EDGE OF THE CATHETER HANDLE IN THE PROTECTING TUBE. REVIEW OF THE DEVICE HISTORY RECORDS CONFIRMED THERE WAS NO ISSUE RELATED TO COMPLAINT DURING MFG PROCESS. A QUALITY IMPROVEMENT PROJECT WAS INITIATED TO ADDRESS THIS ISSUE. DATE REPORT SUBMITTED TO FDA BY MFR: 09/21/2010. DATE THE INITIAL REPORTER PROVIDED THE INFO TO THE MFR: (B)(6) 2010.

Description of Event or Problem · 1

IT WAS REPORTED, THE CATHETER HANDLE LEAKED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COOL PATH CATHETER 7F, 1304-CP-7-25-L NONE OAD ST JUDE MEDICAL, IRVINE 83353 K28309

Patients

Seq Age Sex Outcome Treatment
1 38 YR