FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 1860096
·
Received September 21, 2010
Report
- Report Number
- 1720753-2010-03139
- Event Type
- Malfunction
- Date Received
- September 21, 2010
- Date of Event
- September 3, 2010
- Report Date
- September 21, 2010
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE FAILURE COULD NOT BE DUPLICATED. THE WALL AC POWER OUTLET WAS MEASURED AND FOUND OUTSIDE OF SPECIFICATIONS DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE 980 SYSTEM LEFT MONITOR WAS FLICKERING DURING A CASE. THE 9800 SYSTEM HAD TO BE REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER SYSTEM. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |