FDA Adverse Event
Malfunction
Summary report: N
SYNCHRON CX9 ALX
MDR report key: 1860010
·
Received October 7, 2010
Report
- Report Number
- 2050012-2010-00901
- Event Type
- Malfunction
- Date Received
- October 7, 2010
- Date of Event
- September 16, 2010
- Report Date
- October 5, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJC
- PMA / PMN Number
- K950958
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED: THE FSE REPLACED WASTE CANISTER FLOAT SWITCH, WATER RESERVOIR FLOAT SWITCH, AND THE CX3 WASTE LINE TO EXIT CANISTER.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC., (BCI) REGARDING WASTE LEAK FROM THE HYDRO ON THE SYNCHRON CX9 ALX ANALYZER. NO INJURY WAS REPORTED IN CONNECTION WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON CX9 ALX | CLINICAL CHEMISTRY ANALYZER | JJC | BECKMAN COULTER INC. | CX9 ALX |