FDA Adverse Event Malfunction Summary report: N

SYNCHRON CX9 ALX

MDR report key: 1860010 · Received October 7, 2010

Report

Report Number
2050012-2010-00901
Event Type
Malfunction
Date Received
October 7, 2010
Date of Event
September 16, 2010
Report Date
October 5, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJC
PMA / PMN Number
K950958
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED: THE FSE REPLACED WASTE CANISTER FLOAT SWITCH, WATER RESERVOIR FLOAT SWITCH, AND THE CX3 WASTE LINE TO EXIT CANISTER.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC., (BCI) REGARDING WASTE LEAK FROM THE HYDRO ON THE SYNCHRON CX9 ALX ANALYZER. NO INJURY WAS REPORTED IN CONNECTION WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYNCHRON CX9 ALX CLINICAL CHEMISTRY ANALYZER JJC BECKMAN COULTER INC. CX9 ALX

Patients

Seq Age Sex Outcome Treatment
1