FDA Adverse Event Malfunction Summary report: N

ALARIS SYSTEM

MDR report key: 18594322 · Received January 29, 2024

Report

Report Number
2016493-2024-11257
Event Type
Malfunction
Date Received
January 29, 2024
Date of Event
December 12, 2023
Report Date
January 31, 2024
Manufacturer
CAREFUSION SD
Product Code
FRN
UDI-DI
10885403811043
PMA / PMN Number
K133532
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHARMACIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A FOLLOW UP REPORT WILL BE SUBMITTED WITH INVESTIGATION RESULTS SHOULD THE DEVICE BE REPAIRED OR THE DEVICE/LOGS BE RECEIVED FOR EVALUATION. PER 803.52(F)(11)(III) THE INFORMATION PROVIDED REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. THE COMPLAINANT OR REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO THE MANUFACTURER. H3 OTHER TEXT : DEVICE WAS NOT RETURNED TO MANUFACTURING FACILITY.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION : IMDRF ANNEX B CODE AND MANUFACTURER NARRATIVE. A DEVICE HISTORY RECORD REVIEW IS PERFORMED ON EACH DEVICE REPORTED IN A MDR REPORTABLE EVENT ALONG WITH OTHER METHODS OF INVESTIGATION AS CODED IN SECTION H6 OF THIS MDR REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT CUSTOMER IS NOTING SYRINGE VOLUME DISCREPANCIES WITH NICU ANTIBIOTICS WHILE USING SYRINGE PUMP. THE CUSTOMER IS USING BD 10ML SYRINGES. THE BELOW IS AN EXAMPLE OF GENTAMICIN DRUG. GENTAMICIN¿ 18.92MG IN 9.46ML PER EPIC ORDER. ALARIS READ VOLUMES: 12/12 0849 = 9.07ML. 12/13 0816 = 9.27ML. 12/14 0814 = 9.25ML 12/15 0808 = 9.27ML. 12/16 0846 = 9.46ML. 12/17 0846 = 9.46ML. 12/18 0847 = 9.14ML. THERE WAS PATIENT INVOLVEMENT BUT NO HARM.

Description of Event or Problem · 0

IT WAS REPORTED THAT CUSTOMER IS NOTING SYRINGE VOLUME DISCREPANCIES WITH NICU ANTIBIOTICS WHILE USING SYRINGE PUMP. THE CUSTOMER IS USING BD 10ML SYRINGES. THE BELOW IS AN EXAMPLE OF GENTAMICIN DRUG. GENTAMICIN¿ 18.92MG IN 9.46ML PER EPIC ORDER. ALARIS READ VOLUMES: 12/12 0849 = 9.07ML. 12/13 0816 = 9.27ML. 12/14 0814 = 9.25ML. 12/15 0808 = 9.27ML. 12/16 0846 = 9.46ML. 12/17 0846 = 9.46ML. 12/18 0847 = 9.14ML. THERE WAS PATIENT INVOLVEMENT BUT NO HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1533550 ALARIS SYSTEM PUMP, INFUSION FRN CAREFUSION SD 8110 10885403811043

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown 8015