ALARIS SYSTEM
Report
- Report Number
- 2016493-2024-11256
- Event Type
- Malfunction
- Date Received
- January 29, 2024
- Date of Event
- December 15, 2023
- Report Date
- January 31, 2024
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- UDI-DI
- 10885403811043
- PMA / PMN Number
- K133532
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PHARMACIST
- Health Professional
- Yes
Narratives
A FOLLOW UP REPORT WILL BE SUBMITTED WITH INVESTIGATION RESULTS SHOULD THE DEVICE BE REPAIRED OR THE DEVICE/LOGS BE RECEIVED FOR EVALUATION. PER 803.52(F)(11)(III) THE INFORMATION PROVIDED REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. THE COMPLAINANT OR REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO THE MANUFACTURER. H3 OTHER TEXT : DEVICE WAS NOT RETURNED TO MANUFACTURING FACILITY.
ADDITIONAL INFORMATION: IMDRF ANNEX B CODES AND MANUFACTURER NARRATIVE. A DEVICE HISTORY RECORD REVIEW IS PERFORMED ON EACH DEVICE REPORTED IN A MDR REPORTABLE EVENT ALONG WITH OTHER METHODS OF INVESTIGATION AS CODED IN SECTION H6 OF THIS MDR REPORT.
IT WAS REPORTED THAT CUSTOMER IS NOTING SYRINGE VOLUME DISCREPANCIES WITH NICU ANTIBIOTICS WHILE USING SYRINGE PUMP. THE CUSTOMER IS USING BD 10ML SYRINGES. THE BELOW IS AN EXAMPLE OF GENTAMICIN DRUG. GENTAMICIN¿ 18.92MG IN 9.46ML PER EPIC ORDER. ALARIS READ VOLUMES: 12/12 0849 = 9.07ML. 12/13 0816 = 9.27ML. 12/14 0814 = 9.25ML. 12/15 0808 = 9.27ML. 12/16 0846 = 9.46ML. 12/17 0846 = 9.46ML. 12/18 0847 = 9.14ML. THERE WAS PATIENT INVOLVEMENT BUT NO HARM.
IT WAS REPORTED THAT CUSTOMER IS NOTING SYRINGE VOLUME DISCREPANCIES WITH NICU ANTIBIOTICS WHILE USING SYRINGE PUMP. THE CUSTOMER IS USING BD 10ML SYRINGES. THE BELOW IS AN EXAMPLE OF GENTAMICIN DRUG. GENTAMICIN¿ 18.92MG IN 9.46ML PER EPIC ORDER. ALARIS READ VOLUMES: (B)(6) 0849 = 9.07ML (B)(6) 0816 = 9.27ML (B)(6) 0814 = 9.25ML (B)(6) 0808 = 9.27ML (B)(6) 0846 = 9.46ML (B)(6) 0846 = 9.46ML (B)(6) 0847 = 9.14ML THERE WAS PATIENT INVOLVEMENT BUT NO HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1533549 | ALARIS SYSTEM | PUMP, INFUSION | FRN | CAREFUSION SD | 8110 | 10885403811043 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | 8015 |