CARELINK CONNECT APP ANDROID
Report
- Report Number
- 2032227-2024-116487
- Event Type
- Malfunction
- Date Received
- January 29, 2024
- Date of Event
- January 9, 2024
- Report Date
- April 3, 2024
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- PKU
- PMA / PMN Number
- K151236
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CI
- Reporter Occupation
- 003
Narratives
AN ATTEMPT TO REPRODUCE THE REPORTED EVENT USING THE CP APP WITH 3.2.0[1284] PROD BUILD INSTALLED ON REDMI NOTE 10 (V.12) WITH MMT-1885 PUMP (SOFTWARE VERSION 3.13.9) WAS CONDUCTED AND THE ISSUE WAS REPRODUCED. WE HAVE OBSERVED THAT THE SG GRAPH IS NOT VISIBLE ON THE PATIENT GRAPH SCREEN IN THE CP APP. THE SOFTWARE FAILED TO MEET THE SPECIFIED REQUIREMENTS AND PERFORMANCE EXPECTATIONS, (SRS DOC: 3205046): 3213314, 3213354, 3214374 AGILE DOCS: 10948256DOC. AFTER CONDUCTING A THOROUGH INVESTIGATION, WE FOUND THAT THE DATE AND TIME SETTINGS ON THE PUMP DEVICE AND THE PATIENT DEVICE ARE STILL NOT SYNCHRONIZED, WHICH IS CAUSING THE SG GRAPH TO NOT BE VISIBLE IN THE CP APP. THE ESF NUMBER THAT CORRESPONDS TO THE REPORTED ISSUE IS (B)(4). TO ASSIST WITH THE RESOLUTION OF THE ISSUE, WE PROVIDED THE HELPLINE TEAM WITH THE FOLLOWING WORKAROUND TO ENSURE THAT IT IS ADDRESSED EFFECTIVELY: 1. PLEASE MAKE SURE THAT THE DATE AND TIME ARE THE SAME ON BOTH THE PUMP DEVICE AND THE PATIENT'S DEVICE, BY DOING THIS SG GRAPH WILL BE VISIBLE ON THE CP APP. YOU CAN SET THE DATE AND TIME ON THE PUMP DEVICE USING THE USER GUIDE. THE ISSUE WILL BE RESOLVED IN THE UPCOMING CP APPLICATION RELEASE/S. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. "SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. " MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION DOESN'T SUGGEST THAT THE MALFUNCTION WOULD CAUSE/CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED CARELINK CONNECT ISSUES AND FAILURES IN THE CARELINK CONNECT APPLICATION, WHEN CONNECTING IT TELLS THAT THERE IS NO PATIENT DATA. TROUBLESHOOTING WAS PERFORMED AND UNINSTALLED AND REINSTALLED THE APP BUT THE THE SYNCHRONIZATION WAS VERIFIED IN THE PATIENT'S MINIMED MOBILE BUT IT STILL DOES NOT SHOW THE INFORMATION AND THE ISSUE WAS ESCALATED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL CONTINUE USING THE DEVICE AND THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1648334 | CARELINK CONNECT APP ANDROID | INSULIN PUMP SECONDARY DISPLAY | PKU | MEDTRONIC MINIMED | MMT-6111 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |