FDA Adverse Event Malfunction Summary report: N

SINGLE USE SUCTION VALVE (STERILE)

MDR report key: 18589683 · Received January 26, 2024

Report

Report Number
3003637092-2024-00017
Event Type
Malfunction
Date Received
January 26, 2024
Date of Event
December 27, 2023
Report Date
May 7, 2024
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP. HINODE PLANT
Product Code
EOQ
UDI-DI
14953170152426
PMA / PMN Number
K061313
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED, TO CORRECT THE FDA PRODUCT CODE TO EOQ.

Additional Manufacturer Narrative · 0

THE CLIENT WAS INFORMED BY THE COMPANY REPRESENTATIVE THAT THAT PRACTICE WAS NOT IN ACCORDANCE WITH THE PRODUCT AND IT WAS REQUESTED THAT THE CLIENT DISPOSES THE DEVICE AFTER USE AND THAT A NEW ONE BE PLACED WITH THE ENDOSCOPE BEFORE THE EXAMINATION. THE DEVICE WAS NOT RETURNED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS TO INFORM THAT UPON FURTHER REVIEW, THIS IS NOT A REPORTABLE MALFUNCTION. THE INITIAL MEDWATCH REPORTED THE CLIENT REUSED THE SINGLE-USE PISTONS ON THE BRONCHOSCOPES BY DISASSEMBLING THEM AND PASSING THEM THROUGH THE PERACETIC ACID (PAA) WASHERS. HOWEVER, PER THE LEGAL MANUFACTURER, THIS ISSUE IS NOT LIKELY TO CAUSE OR CONTRIBUTE TO DEATH OR SERIOUS INJURY IF THE MALFUNCTION WERE TO RECUR.

Description of Event or Problem · 0

IT WAS OBSERVED THAT DURING THE DEVICE ESTIMATION, THE SINGLE-USE PISTONS OF THE BRONCHOSCOPES, THE CLIENT REUSED THE SINGLE-USE PISTONS ON THE BRONCHOSCOPES BY DISASSEMBLING THEM AND PASSING THEM THROUGH THE PEREACETIC ACID (PAA) WASHERS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1973472 SINGLE USE SUCTION VALVE (STERILE) SINGLE USE SUCTION VALVE EOQ OLYMPUS MEDICAL SYSTEMS CORP. HINODE PLANT MAJ-209 UNKNOWN 14953170152426
671448 SINGLE USE SUCTION VALVE (STERILE) SINGLE USE SUCTION VALVE EOQ OLYMPUS MEDICAL SYSTEMS CORP. HINODE PLANT MAJ-209 UNKNOWN 14953170152426

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown