FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET

MDR report key: 18588618 · Received January 26, 2024

Report

Report Number
2243072-2024-00069
Event Type
Malfunction
Date Received
January 26, 2024
Date of Event
January 4, 2024
Report Date
February 15, 2024
Manufacturer
BECTON DICKINSON
Product Code
JKA
UDI-DI
00382903672837
PMA / PMN Number
K980414
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY: MATERIAL #: 367283. LOT/BATCH #: 23E08B2. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 50 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION, AND ANOTHER 8 RETENTION SAMPLES BY FUNCTIONAL TESTING, EACH USED TO DRAW 10 VACUTAINER TUBES, AND NO ISSUES WERE OBSERVED RELATING TO SLEEVE LEAKAGE AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE SLEEVE LEAKAGE. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

THE MANUFACTURING LOCATION FOR THIS PRODUCT IS NIPRO. THIS SITE IS AN OEM MANUFACTURING SITE. THEREFORE, BD CORPORATE HEADQUARTERS IN FRANKLIN LAKES, NJ HAS BEEN LISTED IN SECTIONS D.3. AND G.1. AND THE FRANKLIN LAKES FDA REGISTRATION NUMBER HAS BEEN USED FOR THE MANUFACTURE REPORT NUMBER. THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION IS AS FOLLOWS: G.5. PMA / 510(K)#: K991088 H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET THAT CUSTOMER STATES BLOOD DRIPPED FROM END OF SAFETYLOK NEEDLE ONTO FLOOR. NO PATIENT IMPACT.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET THAT CUSTOMER STATES BLOOD DRIPPED FROM END OF SAFETYLOK NEEDLE ONTO FLOOR. NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1121145 BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON DICKINSON 23E08B2 00382903672837

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown