BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET
Report
- Report Number
- 2243072-2024-00069
- Event Type
- Malfunction
- Date Received
- January 26, 2024
- Date of Event
- January 4, 2024
- Report Date
- February 15, 2024
- Manufacturer
- BECTON DICKINSON
- Product Code
- JKA
- UDI-DI
- 00382903672837
- PMA / PMN Number
- K980414
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6 INVESTIGATION SUMMARY: MATERIAL #: 367283. LOT/BATCH #: 23E08B2. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 50 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION, AND ANOTHER 8 RETENTION SAMPLES BY FUNCTIONAL TESTING, EACH USED TO DRAW 10 VACUTAINER TUBES, AND NO ISSUES WERE OBSERVED RELATING TO SLEEVE LEAKAGE AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE SLEEVE LEAKAGE. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
THE MANUFACTURING LOCATION FOR THIS PRODUCT IS NIPRO. THIS SITE IS AN OEM MANUFACTURING SITE. THEREFORE, BD CORPORATE HEADQUARTERS IN FRANKLIN LAKES, NJ HAS BEEN LISTED IN SECTIONS D.3. AND G.1. AND THE FRANKLIN LAKES FDA REGISTRATION NUMBER HAS BEEN USED FOR THE MANUFACTURE REPORT NUMBER. THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION IS AS FOLLOWS: G.5. PMA / 510(K)#: K991088 H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT WHILE USING BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET THAT CUSTOMER STATES BLOOD DRIPPED FROM END OF SAFETYLOK NEEDLE ONTO FLOOR. NO PATIENT IMPACT.
IT WAS REPORTED THAT WHILE USING BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET THAT CUSTOMER STATES BLOOD DRIPPED FROM END OF SAFETYLOK NEEDLE ONTO FLOOR. NO PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1121145 | BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON DICKINSON | 23E08B2 | 00382903672837 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |