FDA Adverse Event Malfunction Summary report: N

OCTARAY MAPPING CATHETER

MDR report key: 18588547 · Received January 26, 2024

Report

Report Number
2029046-2024-00304
Event Type
Malfunction
Date Received
January 26, 2024
Date of Event
January 5, 2024
Report Date
January 26, 2024
Manufacturer
BIOSENSE WEBSTER INC
Product Code
MTD
UDI-DI
10846835021127
PMA / PMN Number
K193237
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIAC ABLATION PROCEDURE WITH A OCTARAY MAPPING CATHETER AND THE SALINE WOULD NOT FLOW OUT FROM THE LUMEN AS USUAL. IT WAS REPORTED AT THE TIME OF PRE-MAPPING, THEY WERE UNABLE TO IRRIGATE PROPERLY. THEY REPLACED THE OCTARAY MAPPING CATHETER RESOLED THE ISSUE. AFTER THAT, THE PROCEDURE WAS COMPLETED WITH NO TROUBLE. THE PROCEDURE WAS COMPLETED WITHOUT PATIENT'S CONSEQUENCE. DEVICE EVALUATION DETAILS: THE DEVICE WAS RETURNED TO BIOSENSE WEBSTER INC (BWI) FOR EVALUATION AND THE EVALUATION HAS BEEN COMPLETED. A VISUAL INSPECTION AND PATENCY TEST OF THE RETURNED DEVICE WERE PERFORMED FOLLOWING BWI PROCEDURES. VISUAL ANALYSIS REVEALED NO DAMAGE OR ANOMALIES ON THE DEVICE. A PATENCY TEST WAS PERFORMED, AND THE DEVICE WAS FLUSHING CORRECTLY. NO IRRIGATION ISSUES WERE OBSERVED. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE BATCH NUMBER, AND NO INTERNAL ACTIONS WERE IDENTIFIED. THE ISSUE REPORTED BY THE CUSTOMER COULD NOT BE REPLICATED DURING THE PRODUCT INVESTIGATION; OTHER ISSUES OR CIRCUMSTANCES MAY HAVE OCCURRED DURING THE USAGE OF THE DEVICE THAT COMPROMISED ITS PERFORMANCE. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING RECOMMENDATIONS: FLUSH THE CATHETER WITH HEPARINIZED SALINE PRIOR TO INSERTION INTO THE BODY. AS PART OF BIOSENSE WEBSTER¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REF. # (B)(4).

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY BIOSENSE WEBSTER INC., OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, BIOSENSE WEBSTER INC., OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. E1. INITIAL REPORTER PHONE: (B)(6). IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REF. # (B)(4).

Additional Manufacturer Narrative · 0

ON 22-FEB-2024, THE BWI PRODUCT ANALYSIS LAB RECEIVED THE COMPLAINT DEVICE FOR EVALUATION. THE PRODUCT ANALYSIS HAS BEGUN BUT IS NOT COMPLETED AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REF. # (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIAC ABLATION PROCEDURE WITH A OCTARAY MAPPING CATHETER AND THE SALINE WOULD NOT FLOW OUT FROM THE LUMEN AS USUAL. IT WAS REPORTED AT THE TIME OF PRE-MAPPING, THEY WERE UNABLE TO IRRIGATE PROPERLY. THEY REPLACED THE OCTARAY MAPPING CATHETER RESOLED THE ISSUE. AFTER THAT, THE PROCEDURE WAS COMPLETED WITH NO TROUBLE. THE PROCEDURE WAS COMPLETED WITHOUT PATIENT'S CONSEQUENCE. ADDITIONAL INFORMATION WAS RECEIVED INDICATING THE ISSUE WAS DISCOVERED WHEN THE SALINE WOULD NOT FLOW OUT FROM THE LUMEN AS USUAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1148407 OCTARAY MAPPING CATHETER CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY MTD BIOSENSE WEBSTER INC 31131881L 10846835021127

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown