FDA Adverse Event Injury Summary report: N

BALL HEADS: MECTACER BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 28 SIZE S - 3.5

MDR report key: 18585791 · Received January 26, 2024

Report

Report Number
3005180920-2024-00002
Event Type
Injury
Date Received
January 26, 2024
Date of Event
January 2, 2024
Report Date
March 28, 2024
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030809125
PMA / PMN Number
K112115
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 05 JANUARY 2024. LOT 2113855: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 20-DEC-2021. EXPIRATION DATE: 2026-DEC-05. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 91 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. OTHER DEVICE INVOLVED: BATCH REVIEW PERFORMED ON 01 MARCH 2024. LINER: VERSAFITCUP DM 01.26.2850MHC DOUBLE MOBILITY HC LINER Ø 50/28 (K092265) LOT 2000167: (B)(4)ITEMS MANUFACTURED AND RELEASED ON 16-APR-2020. EXPIRATION DATE: 2025-MARCH-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 141 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. VISUAL INSPECTION PERFORMED BY R&D PROJECT MANAGER FROM THE RECEIVED PIECES, IT WAS NOT POSSIBLE TO DETERMINE WITH CERTAINTY THE ROOT CAUSE OF THE EVENT; THE BALL HEAD WAS BROKEN IN TWO PARTS, BUT WITHOUT ANY PARTICULAR SIGN OF FAILURE RELATED TO THE PAIN, WHILE THE DM LINER DID NOT SHOW PARTICULAR SIGNS TOO, EXCEPT FOR SOME SCRATCHES MOST PROBABLY OCCURRED FOR THE REVISION AND THE FRICTION WITH THE PIECES OF THE BROKEN HEAD, ALSO BROKEN DURING REMOVAL.

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 05 JANUARY 2024. LOT 2113855: (B)(4) MANUFACTURED AND RELEASED ON 20-DEC-2021. EXPIRATION DATE: 2026-DEC-05. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 91 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW.

Description of Event or Problem · 0

AT ABOUT 1 YEAR AND 4 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN AND THE CAUSE WAS UNKNOWN. DURING THE REVISION SURGERY, A BONE TAMP WAS USED TO REMOVE THE HEAD AND THE DM LINER FROM THE TRUNION. AFTER HITTING ONCE WITH MALLET, THE CERAMIC HEAD BROKE INTO MULTIPLE PIECES WITHIN THE DM LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1973277 BALL HEADS: MECTACER BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 28 SIZE S - 3.5 HIP CERAMIC BALL HEAD LZO MEDACTA INTERNATIONAL SA 2113855 07630030809125

Patients

Seq Age Sex Outcome Treatment
1 48 YR Male Required Intervention