MENTOR MEMORYGEL XTRA BREAST IMPLANT
Report
- Report Number
- 1645337-2024-00964
- Event Type
- Injury
- Date Received
- January 25, 2024
- Date of Event
- December 15, 2023
- Manufacturer
- MENTOR TEXAS
- Product Code
- FTR
- UDI-DI
- 00081317025320
- PMA / PMN Number
- P030053
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: PTOSIS. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
ON FEBRUARY 13, 2024, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. ON FEBRUARY 15, 2024, THE PRODUCT INVESTIGATION WAS COMPLETED. DEVICE INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. DURING VISUAL EVALUATION NO APPARENT DAMAGE OR VISUAL ANOMALIES WERE OBSERVED ON THE SMOOTH HIGH PROFILE XTRA 560CC RETURNED DEVICE. SOME BREAST PTOSIS IS A NORMAL COMPONENT OF THE MATURE BREAST. PTOSIS BECOMES UNDESIRABLE WHEN THE BREAST PARENCHYMA PREDOMINATES BELOW THE AREOLA, DROOPS CONSIDERABLY BELOW THE INFRAMAMMARY FOLD AND THE NIPPLE POINTS DOWNWARD. AUGMENTATION ALONE, WITHOUT CONSIDERATION OF THE PTOSIS CAN PRODUCE A LESS THAN DESIRABLE COSMETIC RESULT KNOWN AS A "ROCK IN SOCK" DEFORMITY OR AN INCREASED PTOTIC APPEARANCE. PTOSIS IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. SINCE NO MALFUNCTION WAS OBSERVED DURING THE INVESTIGATION, NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW.
IT WAS REPORTED THAT A PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH TWO MENTOR MEMORYGEL XTRA BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT SUFFERED RIGHT BREAST CAPSULAR CONTRACTURE (BAKER GRADE III) AND BILATERAL BREAST PTOSIS. THE RIGHT BREAST HAD PAIN AND WAS ELEVATED. AS A RESULT, THE PATIENT UNDERWENT BILATERAL BREAST IMPLANT REMOVAL AND REPLACEMENT SURGERY ON (B)(6) 2024. DURING THE REMOVAL SURGERY, THE RIGHT BREAST IMPLANT WAS DISCOVERED TO HAVE RUPTURED. SUBSEQUENTLY, THE PATIENT'S IMPLANTS WERE REPLACED WITH THE FOLLOWING DEVICES: (RIGHT) 740CC MENTOR MEMORYGEL BOOST BREAST IMPLANT CATALOG: SHPB740 LOT: 9981061 SN: (B)(6) AND (LEFT) 740CC MENTOR MEMORYGEL BOOST BREAST IMPLANT CATALOG: SHPB740 LOT: 9981061 SN: (B)(6). THIS MEDWATCH FORM IS FOR THE LEFT BREAST PROSTHESIS. COMPLICATIONS ON THE RIGHT BREAST/IMPLANT WERE REPORTED UNDER MRN: 1645337-2024-00569.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1898701 | MENTOR MEMORYGEL XTRA BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | MENTOR TEXAS | 7553488 | 00081317025320 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR | Female | Required Intervention |