ENDO GIA
Report
- Report Number
- 2647580-2024-00475
- Event Type
- Malfunction
- Date Received
- January 24, 2024
- Date of Event
- December 19, 2023
- Report Date
- April 3, 2024
- Manufacturer
- US SURGICAL PUERTO RICO
- Product Code
- GDW
- UDI-DI
- 10884523003215
- PMA / PMN Number
- K111825
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
D10: CONCOMITANT PRODUCTS: EGIAUXL, EGIAUXL ENDOGIA ULTRA UNIV XL STAPLER, (LOT#:P2J0021), EGIA60AMT, EGIA60AMT, EGIA 60, ARTIC MED THICK SULU. (LOT#:P2L0208) EGIA60AMT, EGIA60AMT, EGIA 60, ARTIC MED THICK SULU, (LOT#:P2L0208). MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. VISUAL INSPECTION NOTED THAT THE RELOAD WAS PARTIALLY FIRED TO THE CUT LINE AND THE CLAMPING MECHANISM WAS DEFORMED. FUNCTIONALLY, THE RELOAD WAS LOADED INTO A REPRESENTATIVE INSTRUMENT. THE INTERLOCK WAS OVERRIDDEN. THE RELOAD WAS APPLIED TO TEST MEDIA. ALL REMAINING STAPLES WERE PLACED AND TEST MEDIA WAS CLEANLY TRANSECTED. THE RELOAD INTERLOCK WAS TESTED AND FOUND TO FUNCTION PROPERLY. IT WAS REPORTED THAT THE INSTRUMENT DID NOT FIRE. THE REPORTED ISSUE WAS CONFIRMED. THE PRODUCT ANALYSIS NOTED EVIDENCE THAT THE DEVICE WAS NOT USED AS INTENDED. THE ISSUE MAY OCCUR UNDER THE FOLLOWING CONDITIONS: APPLICATION OVER TISSUE THAT IS BEYOND THE RECOMMENDED THICKNESS RANGE OR APPLICATION WITH AN OBSTACLE INCORPORATED IN THE JAWS. IN ANY OF THESE CIRCUMSTANCES, IT WILL BECOME INCREASINGLY DIFFICULT TO ACTUATE THE FIRING HANDLE AND THE INSTRUMENT RETURN KNOBS WILL BE DIFFICULT TO RETRACT. IN ADDITION, STAPLES MAY NOT FORM PROPERLY AND TISSUE MAY NOT BE FULLY TRANSECTED. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. THE INSTRUCTIONS INCLUDED WITH THIS DEVICE PROVIDE THE FOLLOWING GUIDANCE: FAILURE TO COMPLETELY FIRE THE RELOAD WILL RESULT IN AN INCOMPLETE CUT OR INCOMPLETE STAPLE FORMATION, WHICH MAY RESULT IN POOR HEMOSTASIS OR LEAKAGE. ALWAYS INSPECT THE TISSUE THICKNESS AND SELECT AN APPROPRIATE STAPLE SIZE PRIOR TO APPLICATION OF THE ENDO GIA¿ ULTRA UNIVERSAL SHORT, ENDO GIA¿ ULTRA UNIVERSAL OR ENDO GIA¿ ULTRA UNIVERSAL XL STAPLER. OVERLY THICK OR THIN TISSUE MAY RESULT IN UNACCEPTABLE STAPLE FORMATION. WHEN POSITIONING THE STAPLER ON THE APPLICATION SITE, ENSURE THAT NO OBSTRUCTIONS, SUCH AS CLIPS, ARE INCORPORATED INTO THE INSTRUMENT JAWS. FIRING OVER AN OBSTRUCTION MAY RESULT IN INCOMPLETE CUTTING ACTION OR IMPROPERLY FORMED STAPLES. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
ACCORDING TO THE REPORTER, DURING A GASTROPLASTY, THE DEVICE FIRED ONCE AND THEN JAMMED. THE RELOAD WAS CHANGED TWICE AND ONLY HAD ONE TRIP. THE STAPLER BROKE DUE TO FORCE REQUIRED TO FIRE IT. A NEW HANDLE AND RELOAD FROM DIFFERENT BATCH WERE USED TO RESOLVE THE ISSUE. NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1232950 | ENDO GIA | STAPLE, IMPLANTABLE | GDW | US SURGICAL PUERTO RICO | EGIA60AMT | P2L0208 | 10884523003215 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Female | SEE H10. |