BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER
Report
- Report Number
- 9617032-2024-00063
- Event Type
- Malfunction
- Date Received
- January 23, 2024
- Date of Event
- January 3, 2024
- Report Date
- January 8, 2024
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- JKA
- UDI-DI
- 30382903688358
- PMA / PMN Number
- K991088
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FI
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
E1: INITIAL REPORTER PHONE # :(B)(6). H.6. INVESTIGATION SUMMARY: MATERIAL #: 368835. LOT/BATCH #: 3242435. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING, EACH USED TO DRAW 6 VACUTAINER TUBES, AND NO ISSUES WERE OBSERVED RELATING TO HOLDER SEPARATION AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE HOLDER SEPARATION. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED.
IT WAS REPORTED THAT WHILE USING BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER, THE HOLDER CAME AWAY FROM NEEDLE. THIS EVENT OCCURRED 4 TIMES AND NO REPORT OF ADVERSE OR INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1855389 | BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON AND COMPANY (BD) | 3242435 | 30382903688358 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |