BLEASESIRIUS ANESTHESIA SYSTEM
Report
- Report Number
- 3023361-2010-00016
- Event Type
- Malfunction
- Date Received
- June 23, 2010
- Date of Event
- June 14, 2010
- Report Date
- June 21, 2010
- Manufacturer
- SPACELABS HEALTHCARE LTD.
- Product Code
- BSZ
- PMA / PMN Number
- K051629
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE ACTUAL DEVICE WAS EVALUATED AT THE CUSTOMER'S LOCATION AND REPAIRED BY REPLACING THE TOUCHSCREEN ASSEMBLY. A REVIEW OF SIMILAR DEVICE EVALUATIONS, CONDUCTED AT SPACELABS, IDENTIFIED A CONDITION WHERE THE TOUCHSCREEN POWER CONNECTION COULD SHORT TO GROUND CAUSING THE VENTILATOR DISPLAY TO BLANK. WE HAVE CONCLUDED THAT IT IS POSSIBLE THAT OTHER BLEASESIRIUS DEVICES WITH SIMILAR TOUCHSCREEN ASSEMBLIES MAY FAIL IN A SIMILAR MANNER. SPACELABS HAS DECIDED TO IMPLEMENT A FIELD CORRECTION TO PREVENT OTHER CUSTOMERS FROM EXPERIENCING SIMILAR PROBLEMS. SPACELABS IS IN THE PROCESS OF NOTIFYING THE FDA'S (B)(4) DISTRICT OFFICE OF THIS ACTION.
A POLISH HOSPITAL REPORTED THAT WHEN THE CLINICIAN TOUCHED THE BLEASESIRIUS MACHINE TOUCHSCREEN, THE DISPLAY ON THE VENTILATOR PORTION OF THE ANESTHESIA MACHINE WENT BLANK. THE VENTILATOR ITSELF CONTINUED TO FUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BLEASESIRIUS ANESTHESIA SYSTEM | ANESTHESIA GAS-MACHINE | BSZ | SPACELABS HEALTHCARE LTD. | 14200100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |