HAMILTON MEDICAL AG
Report
- Report Number
- 3001421318-2023-26430
- Event Type
- Malfunction
- Date Received
- January 23, 2024
- Date of Event
- October 1, 2020
- Report Date
- January 22, 2024
- Manufacturer
- HAMILTON MEDICAL AG
- Product Code
- CBK
- UDI-DI
- 00730002856789
- PMA / PMN Number
- K181216
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE COMPLAINT HAS BEEN REOPENED AND REVIEWED ACCORDING TO FDA FORM 483 INSPECTIONAL OBSERVATION EMS #2, EOBS2 FROM THE FDA INSPECTION CONDUCTED BETWEEN JULY 17 TO JULY 21, 2022 AT THE EMS AND BONADUZ SITES. A DETAILED INVESTIGATION WAS PERFORMED BY AN EXPERT FROM THE TECHNICAL SERVICE: THIS ISSUE IS DEEMED A REPORTABLE EVENT SINCE THE MALFUNCTION LED TO AN INOPERABLE DEVICE. THE ROOT CAUSE IS A MALFUNCTION OF THE ESM BOARD. WITHIN THIS INVESTIGATION IT HAS BEEN CONFIRMED THAT THE DEVICE FAILED TO MEET ITS SPECIFICATIONS WHEN THE VENTILATOR WAS SWITCHED ON. THERE WAS NO PATIENT OR USER HARM. THE ALLEGATION IN THIS COMPLAINT WAS CONFIRMED TO BE A COMPLAINT. NO FURTHER INVESTIGATION OR CORRECTION WILL BE PERFORMED EXCEPT THOSE MENTIONED ABOVE. IN FUTURE HAMILTON MEDICAL AG WILL REPORT AN EVENT LIKE THIS ISSUE AS IT WILL BE DEEMED A REPORTABLE EVENT.
MACHINE IS SHOWING TECHNICAL EVENT (246010 ; 233006 ; 233004 ; 233003; 233001; 231022; 285001; 285002) MACHINE IS ALSO SHOWING FAN FAILED & ERROR READ HARDWARE COMPONENT EEPROM. VENTILATOR IS ALSO NOT SWITCHING ON IN AC POWER
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 797099 | HAMILTON MEDICAL AG | HAMILTON-C1 | CBK | HAMILTON MEDICAL AG | HAMILTON-C1 | 00730002856789 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |