FDA Adverse Event Injury Summary report: N

ATTAIN STARFIX

MDR report key: 1855220 · Received October 6, 2010

Report

Report Number
2649622-2010-09686
Event Type
Injury
Date Received
October 6, 2010
Date of Event
June 30, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P060039
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD HAD AN APPARENT FRACTURE, AN IMPEDANCE SPIKE THEN RETURNED TO BASELINE AND SPIKED AGAIN. THERE WAS NO CAPTURE AT MAXIMUM OUTPUT. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN STARFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4195 ASKU

Patients

Seq Age Sex Outcome Treatment
1 39 YR Other| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4076 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD