FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1855218 · Received October 6, 2010

Report

Report Number
2649622-2010-09684
Event Type
Injury
Date Received
October 6, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATRIAL LEAD WAS UNDERSENSING WHICH CAUSED THE DIAGNOSTICS TO BE INCORRECT. THE ATRIAL SENSITIVITY WAS REPROGRAMMED. IT WAS FURTHER REPORTED THAT THE PATIENT HAD FALLEN. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 83 YR Required Intervention (B)(4) IMPLANTABLE PULSE GENERATOR| 5092 IMPLANTABLE PACING LEAD