FDA Adverse Event
Malfunction
Summary report: N
ATTAIN ABILITY
MDR report key: 1854922
·
Received October 5, 2010
Report
- Report Number
- 2649622-2010-09214
- Event Type
- Malfunction
- Date Received
- October 5, 2010
- Date of Event
- August 17, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- OJX
- PMA / PMN Number
- P080006
- Removal / Correction Number
- ASKU
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ANALYSIS FOUND NO ANOMALIES. THE FULL LEAD WAS RETURNED FOR ANALYSIS.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING IMPLANT, THE LEFT VENTRICULAR LEAD WAS UNABLE TO BE IMPLANTED INTO A BRANCH. THE LEAD WAS NOT USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATTAIN ABILITY | IMPLANTABLE PACING LEAD | OJX | MEDTRONIC PUERTO RICO, INC. | 4196 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR | Required Intervention |