CONCERTO CRT-D DR
Report
- Report Number
- 6000144-2010-04312
- Event Type
- Injury
- Date Received
- October 5, 2010
- Date of Event
- June 30, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- NIK
- PMA / PMN Number
- P010031/S031
- Removal / Correction Number
- Z-0126-2010
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. CERAMIC CAP - LEAKAGE-UNSPECIFIED
IT WAS REPORTED THAT THERE WAS POSSIBLE PREMATURE BATTERY DEPLETION AND THE DEVICE HAS REACHED ELECTIVE REPLACEMENT INDICATORS. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THERE WAS POSSIBLE PREMATURE BATTERY DEPLETION AND THE DEVICE HAS REACHED ELECTIVE REPLACEMENT INDICATORS. THE DEVICE IS SCHEDULED TO BE EXPLANTED AND REPLACED ON (B)(6) 2010. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONCERTO CRT-D DR | ASKU | NIK | MEDTRONIC MED REL, INC. | C154DWK | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Hospitalization| R | 6949 IMPLANTABLE TACHY LEAD| 4194 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD |