FDA Adverse Event Injury Summary report: N

SIGMA 300 SR

MDR report key: 1854871 · Received October 5, 2010

Report

Report Number
6000094-2010-01785
Event Type
Injury
Date Received
October 5, 2010
Date of Event
August 12, 2010
Manufacturer
MEDTRONIC S.A.
Product Code
DXY
PMA / PMN Number
P980035/S2
Removal / Correction Number
Z-1510/1519-2009
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) PRELIMINARY ANALYSIS REVEALED NO PACING OUTPUT AND NO TELEMETRY. FURTHER TESTING REVEALED THAT THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT IMPLANTED WITH THIS IMPLANTABLE PULSE GENERATOR (IPG) SUFFERED FROM BRADYCARDIA SYMPTOMS WITH LOSS OF OUTPUT AND NO TELEMETRY. ABNORMAL BATTERY DEPLETION WAS SUSPECTED. THE DEVICE WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIGMA 300 SR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC S.A. SSR303 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention