FDA Adverse Event Injury Summary report: N

CAPSURE SP

MDR report key: 1854860 · Received October 5, 2010

Report

Report Number
2649622-2010-09282
Event Type
Injury
Date Received
October 5, 2010
Date of Event
August 30, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P830061/S12
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CORRECTION: CHANGED EVENT DATE AND FACILITY AWARE DATE. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. THE PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED AND REVEALED VARYING RESISTANCE/IMPEDANCE. IMPEDANCE TREND APPEARS UNSTABLE, VARYING ANYWHERE FROM 484 OHMS TO 2509 OHMS FROM (B)(4) 2009 THROUGH (B)(4) 2010.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS HIGH IMPEDANCE AND A LEAD FRACTURE. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE SP ASKU DTB MEDTRONIC PUERTO RICO, INC. 4024 ASKU

Patients

Seq Age Sex Outcome Treatment
1 84 YR Required Intervention (B)(4) IMPLANTABLE PULSE GENERATOR| (B)(4) IMPLANTABLE PULSE GENERATOR