CAPSURE SP
Report
- Report Number
- 2649622-2010-09282
- Event Type
- Injury
- Date Received
- October 5, 2010
- Date of Event
- August 30, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P830061/S12
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CORRECTION: CHANGED EVENT DATE AND FACILITY AWARE DATE. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. THE PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED AND REVEALED VARYING RESISTANCE/IMPEDANCE. IMPEDANCE TREND APPEARS UNSTABLE, VARYING ANYWHERE FROM 484 OHMS TO 2509 OHMS FROM (B)(4) 2009 THROUGH (B)(4) 2010.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THERE WAS HIGH IMPEDANCE AND A LEAD FRACTURE. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSURE SP | ASKU | DTB | MEDTRONIC PUERTO RICO, INC. | 4024 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 84 YR | Required Intervention | (B)(4) IMPLANTABLE PULSE GENERATOR| (B)(4) IMPLANTABLE PULSE GENERATOR |