FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1854851 · Received October 5, 2010

Report

Report Number
2649622-2010-09118
Event Type
Malfunction
Date Received
October 5, 2010
Date of Event
August 19, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE HELIX WAS DISENGAGED FROM HELICAL CHANNEL. BLOOD/BODY FLUID WAS PRESENT ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), AND IN/ON THE HELIX MECHANISM. THE OUTER INSULATION WAS BREACHED CUT, AND THE OUTER TUBING OVERLAY WAS BREACHED CUT.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER THE HELIX WAS DEPLOYED 4 TO 5 TIMES, IT COULD NO LONGER BE EXTENDED. THE LEAD WAS NOT IMPLANTED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE ASKU LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 52 YR Other| R