FDA Adverse Event Injury Summary report: N

MAXIMO II CRT-D

MDR report key: 1854665 · Received October 5, 2010

Report

Report Number
6000144-2010-04305
Event Type
Injury
Date Received
October 5, 2010
Date of Event
August 16, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S084
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) : THE ACTUAL LONGEVITY IS < 80% OF 99.9% LONGEVITY LIMIT. THE DEVICE WAS FULLY FUNCTIONAL WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

ASKU

Description of Event or Problem · 1

IT WAS REPORTED THAT DEVICE WAS ABRUPTLY AT RRT (RECOMMENDED REPLACEMENT TIME) WITH NO PATIENT THERAPIES. THE OUTPUTS ON ALL THREE LEADS ARE LOW. IT WAS FURTHER REPORTED THE DEVICE WAS AT RRT FASTER THAN EXPECTED. THE DEVICE WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAXIMO II CRT-D ASKU NIK MEDTRONIC MED REL, INC. D284TRK ASKU

Patients

Seq Age Sex Outcome Treatment
1 72 YR Hospitalization| O| R 4193 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD| 6932 IMPLANTABLE TACHY LEAD