VIRTUOSO DR
Report
- Report Number
- 6000144-2010-04299
- Event Type
- Injury
- Date Received
- October 5, 2010
- Date of Event
- August 9, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S062
- Removal / Correction Number
- Z-0128-2010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.
IT WAS REPORTED BY THE PATIENT'S WIFE THAT HER HUSBAND'S DEVICE BATTERY WAS LOW AND NEEDED TO BE REPLACED IN THE NEXT MONTH. SHE STATED THE DEVICE WAS CHECKED A FEW MONTHS AGO, AND NOTHING WAS MENTIONED ABOUT THE BATTERY BEING CLOSE TO REPLACEMENT. THE DEVICE WAS LATER EXPLANTED AND REPLACED. IT WAS RETURNED WITH A REPORT OF PREMATURE BATTERY DEPLETION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VIRTUOSO DR | ASKU | LWS | MEDTRONIC MED REL, INC. | D154AWG | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Required Intervention | 4076 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD| 6949 IMPLANTABLE TACHY LEAD| 4076 IMPLANTABLE PACING LEAD |