FDA Adverse Event
Malfunction
Summary report: N
C-ARM
MDR report key: 1854622
·
Received September 9, 2010
Report
- Report Number
- 1720753-2010-02982
- Event Type
- Malfunction
- Date Received
- September 9, 2010
- Date of Event
- May 27, 2009
- Report Date
- September 9, 2010
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE CONFIGURATION WAS UPDATED AND THE ANALOG INTERFACE PRINTER CIRCUIT BOARD WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE SYSTEM LOST PATIENT DATA. NO REPORT OF PATIENT INJURY.