ARCHITECT HAVAB IGM REAGENT
Report
- Report Number
- 2623532-2010-00073
- Event Type
- Malfunction
- Date Received
- October 5, 2010
- Date of Event
- June 16, 2010
- Report Date
- June 16, 2010
- Manufacturer
- ABBOTT DIAGNOSTICS INT'L, LTD
- Product Code
- LOL
- PMA / PMN Number
- N/A
- Removal / Correction Number
- 2623532-1/4/10-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4): CROSS CONTAMINATION WAS IDENTIFIED AS A POTENTIAL CAUSE. CONCLUSION: INSUFFICIENT ACTIVE ANTIGEN IS BEING COATED ON THE MICROPARTICLE WITH THE CURRENT MICROPARTICLE MANUFACTURING PROCESS. AN ADVERSE TREND IN US CUSTOMER COMPLAINTS FOR ARCHITECT HAVAB-IGM ASSAY (LIST NUMBER 6L21) REGARDING HIGHER THAN USUAL (B)(6) RESULTS RATE WAS DETECTED ON (B)(6) 2009. THE INVESTIGATION REVEALED THE MOST PROBABLE CAUSE FOR THE INCREASE RATE OF (B)(6) RESULTS IS THE HAV ANTIGEN CODE 26286 WHICH HAS A REDUCED POTENCY THAT AFFECTS THE ARCHITECT HAVAB-IGM PERFORMANCE. INSUFFICIENT ACTIVE ANTIGEN IS BEING COATED ON THE MICROPARTICLE WITHIN THE CURRENT MICROPARTICLE MANUFACTURING PROCESS RESULTING IN A PERFORMANCE THAT IS CHARACTERIZED BY LOWER CALIBRATOR 1 RELATIVE LIGHT UNITS (RLU) VALUES. AS A RESULT, THE SIGNAL TO NOISE RATIO IS SHIFTED SPECIALLY IN THE (B)(6) SAMPLES LEAVING THE ASSAY PRONE FOR (B)(6) RESULTS AND INCREASES ARCH HAVAB-IGM ASSAY SUSCEPTIBILITY TO REAGENT CROSS CONTAMINATION AND TEST SYSTEM VARIABILITY. AS A PREVENTIVE MEASURE TO PROTECT THE INTEGRITY OF TEST RESULTS, CUSTOMERS WERE INSTRUCTED BY A PRODUCT INFORMATION LETTER TO FOLLOW AN ALTERNATIVE PROCESSING METHOD BY SEGREGATING ALL HAVAB-IGM SAMPLES.
THE CUSTOMER WAS CONCERNED ABOUT THE INCREASE IN THE NUMBER OF (B)(6) ARCHITECT HAVAB-IGM SAMPLES AFTER IMPLEMENTING THE INSTRUCTIONS IN FA30DEC2009. CURRENT INFORMATION INDICATES ELEVATED (B)(6) PATIENT RESULTS, OR ELEVATED OUT OF RANGE HIGH (B)(6) QUALITY CONTROL RESULTS FOR ARCHITECT HAVAB-IGM MAY OCCUR WHEN THE ASSAY IS RUN DIRECTLY FOLLOWING THE ARCHITECT AUSAB ASSAY. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARCHITECT HAVAB IGM REAGENT | FOR THE DETECTION OF IGM ANTIBODIES TO HEPATITIS A VIRUS | LOL | ABBOTT DIAGNOSTICS INT'L, LTD | 82584Q102 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ARCH I2000SR, LN 3M74-01, SN (B)(4) |