FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 1854431 · Received September 30, 2010

Report

Report Number
1811755-2010-01286
Event Type
Malfunction
Date Received
September 30, 2010
Date of Event
September 15, 2010
Report Date
September 15, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HANDPIECE WAS RECEIVED AT THE MANUFACTURER FOR EVALUATION, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS DUPLICATED. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS PROBLEMS WITH THE DRIVESHAFT, ROTOR, AND SPINDLE HOUSING, WHICH WERE EACH REPLACED ALONG WITH OTHER COMPONENTS. SERVICE DID A CLEAN LUBE, AND ADJUST TO THE DEVICE, AND IT WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HANDPIECE BEGAN OVERHEATING DURING AN EXTRACTION. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER DEVICE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE IMPACTION DRILL SURGICAL INSTRUMENT MOTORS AND ACCESSORIES/ATTAC DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK