FDA Adverse Event Malfunction Summary report: N

GATEWAY BALLOON CATHETER

MDR report key: 1854392 · Received October 5, 2010

Report

Report Number
2939204-2010-00973
Event Type
Malfunction
Date Received
October 5, 2010
Date of Event
September 16, 2010
Report Date
September 21, 2010
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
GBA
PMA / PMN Number
H050001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATION. DEVICE ANALYSIS REVEALED A KINK ADJACENT TO THE EDGE OF THE HUB AND A SECOND KINK 32MM FROM THE EDGE OF THE HUB. MICROSCOPIC INSPECTION REVEALED A PINHOLE IN THE OUTER SHAFT AT THE LOCATION OF THE KINKS. MAGNIFIED INSPECTION OF THE PINHOLE REVEALED NO INDICATION OF ANY CONTRIBUTION FROM MATERIAL OR MANUFACTURING DEFICIENCIES. REVIEW OF THE INFORMATION AVAILABLE FOR THE REPORTED EVENT WAS UNABLE TO DETERMINE AN EXACT ROOT CAUSE. HOWEVER, BASED ON ANALYSIS RESULTS, IT IS POSSIBLE THAT THE DEVICE KINKED DURING HANDLING; WEAKENING THE SHAFT MATERIAL AND RESULTING IN A PINHOLE BURST WHEN PRESSURE WAS APPLIED DURING PROCEDURE. THEREFORE, A ROOT CAUSE OF OPERATIONAL CONTEXT HAS BEEN ASSIGNED TO THIS INVESTIGATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BALLOON CATHETER LEAKED CONTRAST MEDIA AND DEFLATED UPON DILATING. THE PHYSICIAN REMOVED THE BALLOON CATHETER AND REPLACED IT WITH A SIMILAR DEVICE. THERE WAS NO REPORTED CLINICAL CONSEQUENCE TO THE PATIENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BALLOON CATHETER LEAKED CONTRAST MEDIA AND DEFLATED UPON DILATING. THE PHYSICIAN REMOVED THE BALLOON CATHETER AND REPLACED IT WITH A SIMILAR DEVICE. THERE WAS NO REPORTED CLINICAL CONSEQUENCE TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GATEWAY BALLOON CATHETER CATHETER, BALLOON TYPE GBA BOSTON SCIENTIFIC - MAPLE GROVE M0032072209150 13511896

Patients

Seq Age Sex Outcome Treatment
1