FDA Adverse Event Malfunction Summary report: N

CORE MICRO DRILL

MDR report key: 1854390 · Received September 30, 2010

Report

Report Number
1811755-2010-01274
Event Type
Malfunction
Date Received
September 30, 2010
Date of Event
September 7, 2010
Report Date
September 7, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HANDPIECE HAS BEEN REC'D AT THE MFR FOR INVESTIGATION. ACCORDING TO THE INVESTIGATION DETAILS, THE DEVICE REGISTERED HOT DURING TESTING. THE CONTACT PINS IN THE MOTOR CARTRIDGE APPEARED BURNT AND THE ROTOR ASSEMBLY WAS JAMMED IN THE MOTOR ASSEMBLY. THE MOTOR CARTRIDGE AND ROTOR ASSEMBLY WILL BE REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HANDPIECE BEGAN OVERHEATING DURING A SURGICAL PROCEDURE. THERE WAS NO REPORTED PT OR USER INJURY, AND THE CASE WAS COMPLETED SUCCESSFULLY WITH ANOTHER DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE MICRO DRILL POWER SIMPLE CRANIAL DRILLS, BURRS, TREPHINES, A DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK