FDA Adverse Event
Malfunction
Summary report: N
PUREPOINT
MDR report key: 1854334
·
Received September 30, 2010
Report
- Report Number
- 2028159-2010-01867
- Event Type
- Malfunction
- Date Received
- September 30, 2010
- Date of Event
- August 31, 2010
- Report Date
- August 31, 2010
- Manufacturer
- ALCON - IRVINE TECHNOLOGY CENTER
- Product Code
- HQF
- PMA / PMN Number
- K062624
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE COMPANY SERVICE REP EXAMINED THE SYSTEM AND CONFIRMED THE PROBLEM REPORTED. THE PCB WAS REPLACED AND THE CABLES WERE RESEATED. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. (B)(4).
Description of Event or Problem · 1
THE NURSE REPORTED THE SYSTEM WON'T GO ABOVE 10 AND FREEZES WITH NO FIRING TONE. THE SYSTEM WAS SWITCHED OUT TO COMPLETE THE CASE. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUREPOINT | LASER, OPHTHALMIC | HQF | ALCON - IRVINE TECHNOLOGY CENTER | PUREPOINT | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |