FDA Adverse Event Malfunction Summary report: N

PUREPOINT

MDR report key: 1854334 · Received September 30, 2010

Report

Report Number
2028159-2010-01867
Event Type
Malfunction
Date Received
September 30, 2010
Date of Event
August 31, 2010
Report Date
August 31, 2010
Manufacturer
ALCON - IRVINE TECHNOLOGY CENTER
Product Code
HQF
PMA / PMN Number
K062624
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY SERVICE REP EXAMINED THE SYSTEM AND CONFIRMED THE PROBLEM REPORTED. THE PCB WAS REPLACED AND THE CABLES WERE RESEATED. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. (B)(4).

Description of Event or Problem · 1

THE NURSE REPORTED THE SYSTEM WON'T GO ABOVE 10 AND FREEZES WITH NO FIRING TONE. THE SYSTEM WAS SWITCHED OUT TO COMPLETE THE CASE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUREPOINT LASER, OPHTHALMIC HQF ALCON - IRVINE TECHNOLOGY CENTER PUREPOINT NA

Patients

Seq Age Sex Outcome Treatment
1 UNK