FDA Adverse Event Malfunction Summary report: N

7900

MDR report key: 1854155 · Received September 30, 2010

Report

Report Number
9680959-2010-00355
Event Type
Malfunction
Date Received
September 30, 2010
Date of Event
July 30, 2010
Report Date
September 30, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE EMERGENCY STOP BUTTON MECHANISM WAS REPAIRED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM FAILED TO BOOT UP FOLLOWING AN ACTIVATION OF THE FAST STOP. NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 7900 FLUOROSCOPY XRAY JAA GE OEC MEDICAL SYSTEMS GMBH 7900

Patients

Seq Age Sex Outcome Treatment
1