FDA Adverse Event
Malfunction
Summary report: N
7900
MDR report key: 1854155
·
Received September 30, 2010
Report
- Report Number
- 9680959-2010-00355
- Event Type
- Malfunction
- Date Received
- September 30, 2010
- Date of Event
- July 30, 2010
- Report Date
- September 30, 2010
- Manufacturer
- GE OEC MEDICAL SYSTEMS GMBH
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE EMERGENCY STOP BUTTON MECHANISM WAS REPAIRED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE SYSTEM FAILED TO BOOT UP FOLLOWING AN ACTIVATION OF THE FAST STOP. NO REPORT OF PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 7900 | FLUOROSCOPY XRAY | JAA | GE OEC MEDICAL SYSTEMS GMBH | 7900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |