FDA Adverse Event Malfunction Summary report: N

COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL

MDR report key: 1854095 · Received September 2, 2010

Report

Report Number
2023826-2010-00864
Event Type
Malfunction
Date Received
September 2, 2010
Date of Event
July 15, 2010
Report Date
August 9, 2010
Manufacturer
STAAR SURGICAL COMPANY
Product Code
HQL
PMA / PMN Number
P990013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL: METHOD - LENS WORK ORDER SEARCH. RESULTS - VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND A PIECE OF THE LENS OPTIC AND ONE HAPTIC TORN OFF AND MISSING. A PIECE OF THE OTHER HAPTIC WAS TORN OFF AND MISSING. THE LENS WAS RETURNED DRY AND THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. A LENS WORK ORDER SEARCH WAS PERFORMED AND NO SIMILAR COMPLAINTS WERE FOUND WITHIN THE WORK ORDER. CONCLUSIONS: BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH AND THE EVAL OF THE RETURNED PRODUCT, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. (B)(4).

Description of Event or Problem · 1

THE REPORTER INDICATED A CC4204A COLLAMER ASPHERIC SINGLE PIECE LENS BECAME STUCK AND SPLIT IN THE CARTRIDGE. ADDITIONAL INFO HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING. IF ADDITIONAL INFO IS RECEIVED, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL INTRAOCULAR LENS HQL STAAR SURGICAL COMPANY CC4204A NA

Patients

Seq Age Sex Outcome Treatment
1 79 YR CARTRIDGE: MODEL AND LOT NUMBER UNK| INJECTOR: MODEL AND LOT NUMBER UNK