COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL
Report
- Report Number
- 2023826-2010-00864
- Event Type
- Malfunction
- Date Received
- September 2, 2010
- Date of Event
- July 15, 2010
- Report Date
- August 9, 2010
- Manufacturer
- STAAR SURGICAL COMPANY
- Product Code
- HQL
- PMA / PMN Number
- P990013
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- NURSE
Narratives
(B)(4). EVAL: METHOD - LENS WORK ORDER SEARCH. RESULTS - VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND A PIECE OF THE LENS OPTIC AND ONE HAPTIC TORN OFF AND MISSING. A PIECE OF THE OTHER HAPTIC WAS TORN OFF AND MISSING. THE LENS WAS RETURNED DRY AND THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. A LENS WORK ORDER SEARCH WAS PERFORMED AND NO SIMILAR COMPLAINTS WERE FOUND WITHIN THE WORK ORDER. CONCLUSIONS: BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH AND THE EVAL OF THE RETURNED PRODUCT, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. (B)(4).
THE REPORTER INDICATED A CC4204A COLLAMER ASPHERIC SINGLE PIECE LENS BECAME STUCK AND SPLIT IN THE CARTRIDGE. ADDITIONAL INFO HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING. IF ADDITIONAL INFO IS RECEIVED, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL | INTRAOCULAR LENS | HQL | STAAR SURGICAL COMPANY | CC4204A | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | CARTRIDGE: MODEL AND LOT NUMBER UNK| INJECTOR: MODEL AND LOT NUMBER UNK |