FDA Adverse Event Malfunction Summary report: N

C-ARM

MDR report key: 1854058 · Received September 2, 2010

Report

Report Number
1720753-2010-02910
Event Type
Malfunction
Date Received
September 2, 2010
Date of Event
August 18, 2010
Report Date
September 2, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE GENERATOR INTERFACE BOARD WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE DEVICE DISPLAYED A COMMUNICATION FAILURE ERROR MESSAGE AND LOCKED UP. NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 C-ARM FLUOROSCOPY XRAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1