FDA Adverse Event Injury Summary report: N

VENTAK PRIZM

MDR report key: 1853987 · Received October 5, 2010

Report

Report Number
2124215-2010-15938
Event Type
Injury
Date Received
October 5, 2010
Date of Event
July 19, 2010
Report Date
July 19, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ERI AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE ERI CHARGE TIME LIMIT OF 18 SECONDS. THE DEVICE PASSED A SERIES OF DIAGNOSTIC TESTS TO ASSESS THE PERFORMANCE OF DEFIBRILLATION, PACING, SENSING, HIGH VOLTAGE SHOCKING, AND DIAGNOSTIC RECORDING FUNCTIONS. THE DEVICE'S MONITORING VOLTAGE WAS NORMAL BASED ON THERAPY USE AND PROGRAMMED SETTINGS. CHARGE TIMES IN EXCESS OF THE 18 SECOND ERI CHARGE TIME LIMIT TRIGGERED ERI EARLIER THAN EXPECTED. THIS WAS DUE TO A BUILD-UP OF INTERNAL BATTERY IMPEDANCE, PREVENTING THE DEVICE FROM MEETING ITS EXPECTED LONGEVITY PER DEVICE LABELING. THIS ISSUE IS DISCUSSED IN OUR Q2 2010 CRM PRODUCT PERFORMANCE REPORT. THIS EVENT WILL BE UPDATED IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER APPROXIMATELY 54 MONTHS OF IMPLANT, WITH A BATTERY MONITORING VOLTAGE OF 2.71 V, AND A CHARGE TIME OF 19.5 SECONDS. THE DEVICE WAS SUCCESSFULLY REPLACED WITH ANOTHER BOSTON SCIENTIFIC DEVICE, AND RETURNED FOR ANALYSIS. NO ADVERSE PATIENT EFFECTS WERE REPORTED IN THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VENTAK PRIZM IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND 1852

Patients

Seq Age Sex Outcome Treatment
1 62 YR Required Intervention E102| 1857| 1852| 1749| 1555| 0158| 0074