FDA Adverse Event Malfunction Summary report: N

ANGLED MEDIUM SABER ATTACHMENT

MDR report key: 1853881 · Received September 7, 2010

Report

Report Number
1811755-2010-01061
Event Type
Malfunction
Date Received
September 7, 2010
Date of Event
August 3, 2010
Report Date
August 12, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SF
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DRILL ATTACHMENT WAS RETURNED TO THE MFR AND THE COMPLAINT WAS CONFIRMED. BASED ON THE RESULTS OF THE QUALITY INVESTIGATION, THE NOSE BEARINGS WERE SEIZED,DUE TO EXTENSIVE CORROSION. CORROSION WITHIN BEARINGS WILL CAUSE EXCESSIVE FRICTION, WHICH CAN LEAD TO HEAT BEING GENERATED. IMPROPER OR INADEQUATE CLEANING/STERILIZATION TECHNIQUES COULD CONTRIBUTE TO CORROSION OF INTERNAL COMPONENTS WITHIN THIS DEVICE. THE DRILL ATTACHMENT COULD NOT BE REPAIRED AND WAS DISCARDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE DRILL ATTACHMENT HEATED UP. BACKUP EQUIPMENT WAS USED TO COMPLETE THE PROCEDURE, WITHOUT CAUSING A CLINICALLY SIGNIFICANT DELAY. NO PT OR USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANGLED MEDIUM SABER ATTACHMENT DRILLS, SURGICAL, ENT (ELECTRIC OR PNEUMATIC), INC ERL STRYKER INSTRUMENTS KALAMAZOO 09222

Patients

Seq Age Sex Outcome Treatment
1 UNK