FDA Adverse Event Malfunction Summary report: N

CONTAK RENEWAL

MDR report key: 1853489 · Received October 5, 2010

Report

Report Number
2124215-2010-14456
Event Type
Malfunction
Date Received
October 5, 2010
Date of Event
May 19, 2010
Report Date
July 16, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Removal / Correction Number
Z-0047-2008 TO Z-0053-20
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT IN OUR POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL INSPECTION CONFIRMED THE CLINICAL OBSERVATION OF A SEPARATED HEADER. ALTHOUGH EVIDENCE INDICATES EXTERNAL STRESS MAY HAVE BEEN A FACTOR IN CAUSING THE HEADER TO BECOME LOOSE, THE ROOT CAUSE OF THE ABNORMALITY WAS ISOLATED TO INSUFFICIENT BONDING BETWEEN THE MEDICAL ADHESIVE AND THE TITANIUM CASING. MANUFACTURING ENHANCEMENTS WERE IMPLEMENTED IN 2003 TO MAKE THE BOND MORE ROBUST. THIS ISSUE IS DISCUSSED IN BOSTON SCIENTIFIC'S PRODUCT PERFORMANCE REPORT.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS EXPLANTED DUE TO NORMAL BATTERY DEPLETION. UPON EXPLANT THE PHYSICIAN ADMITTED TO USING FORCE CAUSING THE HEADER TO POP BUT IT HAD NOT SEPARATED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H170

Patients

Seq Age Sex Outcome Treatment
1 79 YR A155| N119| H170| 0148| 4543| 4087