MDR report key: 1853486 · Received October 5, 2010

Report

Report Number
2124215-2010-14812
Event Type
Malfunction
Date Received
October 5, 2010
Date of Event
July 1, 2010
Report Date
July 20, 2010
Manufacturer
EXTERNAL MANUFACTURER
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE COMMUNICATOR WAS RETURNED FOR ANALYSIS. THE POST MARKET QUALITY ASSURANCE LAB CONFIRMED THE REPORTED ALLEGATION. THE POWER SUPPLY DISPLAYED SIGNS OF MELTING ALONG THE TOP EDGE NEAR THE STRAIN RELIEF AND AN AUDIBLE RINGING COULD BE HEARD DURING TESTING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT STATED THE POWER SUPPLY WAS HOT WITH A BURNING SMELL AND AN AUDIBLE SOUND COMING FROM IT. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COMMUNICATOR ACCESSORY LWS EXTERNAL MANUFACTURER 6476

Patients

Seq Age Sex Outcome Treatment
1 57 YR (B)(4)| N119| 6948| 4193| 4592