THERMATRX OFFICE THERMO THERAPY SYSTEM
Report
- Report Number
- 2183959-2010-00353
- Event Type
- Malfunction
- Date Received
- September 28, 2010
- Date of Event
- July 30, 2010
- Report Date
- August 27, 2010
- Manufacturer
- AMERICAN MEDICAL SYSTEMS, INC.
- Product Code
- MEQ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE CONSOLE WAS RECEIVED AND EVALUATED ON (B)(4) 2010. IT WAS DECONTAMINATED AND ITS EXTERIOR WAS VISUALLY INSPECTED. A NON-RELATED RETURNED GOODS PT CABLE, S/N (B)(4), WAS USED DURING THE FUNCTIONAL TESTING OF THE CONSOLE. THE CONSOLE WAS TESTED 5 TIMES OVER THE TWO-DAY PERIOD (B)(4) 2010 AND (B)(4) 2010. IN ALL FIVE CASES IT PERFORMED IN A ROUTINE MANNER WITH NO ISSUES OR READINGS OUTSIDE OF ESTABLISHED PARAMETERS. SYS SELF-CHECK, CALIBRATION AND OPERATION WERE NORMAL. THE STORED PT FILES WERE DOWNLOADED AND PRINTED OUT FOR INSPECTION. AMS IS STILL IN THE PROCESS OF REVIEWING MORE DATA AND FURTHER TESTS. AS SOON AS AMS IS AWARE OF ADDITIONAL INFO AND ABLE TO DRAW A CONCLUSION WE WILL FILE A FOLLOW-UP REPORT.
RELATED TO MFR REPORT# 2183959-2010-00354. ON (B)(6) 2010 PT HAD A THERMATRX THERMAL TREATMENT. DOCTOR CLAIMED INCONSISTENT READINGS ON THE TMX3000 UNIT AND NOW CLAIMS THE PT IS INCONTINENT. THERMATRX SHOW A MAX TEMP SENSOR READING OF 66 DEGREES HOWEVER THE GRAFT SHOWS 44 MAX. SECOND EXAMPLE IS THIS UNIT SHOWS A TREATMENT TIME OF 80 MINS WITH ONLY A 40 MIN TREATMENT TIME. THE GRAPH WATTAGE READING IS OFF THE CHART.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THERMATRX OFFICE THERMO THERAPY SYSTEM | TRANSURETHRAL MICROWAVE THERMOTHERAPY SYSTEM | MEQ | AMERICAN MEDICAL SYSTEMS, INC. | TMX-3000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI | Disability |